• Increase font size
  • Default font size
  • Decrease font size

Atomic Fluorescence Spectrometers

asf200t

AFS 200T Atomic Fluorescence Spectrometer

Product description











Performance characteristics




Applications

Hydride-Atomic Fluorescence Spectrometry (HG-AFS) is a new hyphenated analysis technology, developed from the perfect combination of hydride generation technology and non-dispersive atomic fluorescence spectrometry system. It can test elements which can form gaseous hydride such as AS, Sb, Bi, Se, Te, Pb, Sn, Ge and elements which can form volatile gaseous components such as Hg, Cd, Zn.
HG-AFS has obtained great evaluation from analyzers domestic and abroad with its advantages such as simple structure, high sensitivity, little GC interference, wide linear range, fast analysis speed. AFS200T Double-channel Atomic Fluorescence Spectrometer is elaborated by Skyray, using national patented technology and having own intellectual property right. Its well performance and complete functions can meet users’ comprehensive sample testing requirements.

Non-dispersive optical system of simple structure can measure two elements at most.
Automatic controls such as lamp current, negative high voltage, peristaltic pump RPM; USB2.0.
With function of Ar gas flow auto control, the instrument can cut off Ar gas when stop working, convenient for users’ operation and saving gas source.
In-built device for flame observation is convenient for checking instrument running status.

Geology, environment, metallurgy, food, biology, medicine, agricultural plant sample, light industry, cosmetics and so on.
Specifications

Model

AFS 200T, Atomic Fluorescence Spectrometer

Measurable range

11 elements such as As, Sb, Bi, Hg, Se, Te, Pb, Sn, Ge, Zn, Cd

Typical detection limit As, Sb ≤ 0,01 ng/ml;  Hg ≤ 0,001 ng/ml
Accuracy Accuracy for typical elements ≤ 1%
Linear range

More than three orders of magnitude

Standard configurations Low temperature atomizer
Two peristaltic pumps
Electric circuit control system
Optical multiplier
CCD of high resolution
Two mass flow controllers
Atomizer height fine-tuning device
Optical system

This specification can be changed without prior notice.