EDXRF Spectrometers |
|
|
|
|
|
EDX3600B XRF Spectrometer |
|
|
Product description |
EDX3600B X-ray Fluorescence spectrometer uses XRF technology for rapid and accurate elemental analysis of cement and steel. The technology features low-energy X-rays with good excitation results of light elements such as Si, S, Na and Mg. Plus, the test time is short and test efficiency is greatly improved. The instrument is equipped with UHRD detector as well, which leads to good energy linearity, energy resolution, spectrum property and high peak-background ratio. Due to automatic spectrum stabilizing device, the instrument is in great consistency. As spectrum decoupling technique is used, the original spectrum can be easily decoupled, and the measured analytical precision of light elements of Si, S, Al, etc is the same. Owing to multi-parameter linear regression method, the absorption and enhancement effects between elements are significantly reduced.
Deliver professional full-element analysis of cement, steel, minerals, plating thickness and hazardous elements (RoHS). |
| Applications |
Mainly used for full-element analysis in cement and mineral industry |
| Specifications | |
|
Model |
EDX 3600B, proffesional EDXRF Spectrometer for full-element analysis |
|
Detector |
Si-PIN detector electrically cooled, UHRD technology |
|
Excitation source |
Tungsten X-ray tube, |
| Tube current | 50-1000 µA |
|
Tube voltage |
5-50 kV |
| Test time |
60-200 s |
| Measured elements |
From Na to U, improved precision for light elements |
| Analysis range | 1 ppm - 99,99% |
| Measurement precision |
0,05% |
| Simultaneous analysis | 24 elements |
| Forms of samples | solid, powder and liquid |
| Measured layers | 11 layers, up to 0,005 µm thick |
|
Sample positioning |
High resolution CCD camera 1,4 mln pix. |
| Collimators | Collimators and filters can be switched automatically for different samples. |
|
Power |
110/220 VAC +/- 5V (recommended stabilized voltage power supply) |
|
Energy resolution |
150 +/- 5 eV |
| Ambient temperature | +15 °C to + 30 °C |
| Ambient humidity | 35% - 70% |
| Sample chamber dimensions | 320 mm dia. x180 mm h. |
| Weight | 75 kg |
| Standard configurations | Single vacuum sample chamber, high resolution CCD, Si-PIN UHRD detector, light path enhancement system, Signal-to-Noise Enhancer (SNE), Automatic collimator and filter switch, Precise mobile platform, Enhanced metal sensitivity analyzer, full-element analytical software. |
|
|