Atomic Absorption Spectrometers

aas6000-b

AAS 6000 Atomic Absorption Spectrometer

Product description

























Performance characteristics

AAS6000 Series are Single Beam Atomic Absorption Spectrometers controlled and data processed by external computer and internal CPU chips. AAS6000 can be used to measure the concentration of macro, micro and trace metal elements and half-metal elements in various kinds of substances.
Atomic absorption spectrometry is a technique exploiting the fact that atoms at ground state may absorb the radiation of characteristic wavelength. As is known, the atoms are usually at ground state. For each element, the amount of energy required for an atom to transit from ground state to excited state is specific, and in general, is called characteristic wavelength. The Atomic absorption spectrometry typically makes use of hollow cathode lamp as the light source to emit the characteristic wavelength of a particular element. When the light passes through the atomic vapor, it is absorbed by the vapor. The concentration of the element can then be determined through the calculation of the absorption rate.
AAS6000 is provided with three reading methods Continuum, Retention and Peak Height for measuring absorbency, density and emission intensity. It has three signal modes: Atomic Absorption, Background Absorption and Background Correction Absorption, The reading time ranges from 0.5 s to 99s. AAS6000 include 8 computer controllable hollow cathode lamp holders. All working conditions of the instrument such as lamp number, lamp current, negative high voltage, working wavelength, slit, burner location, ignition/extinction and adjustment of burning ratio can be set by keyboard input. The functions of the instrument include automatic gain/ automatic zero, background correction, automatic energy balance, automatic peak location/wavelength scan, and automatic peak location on basis of the retrieval of peak values. All the readings, measurement results, calibration curves and operation conditions can be saved or printed out.

C2H2 monitoring; flame measurement; all reflective, aberration corrected
Fully automated instrument, total automatic operation
8 element lamps, time saved for changing lamps
Gas path safety measures, safer operation
USB interface allows you to operate the instrument conveniently and rapidly.
230nm grating blazed wavelength, multiplied sensitivity in the Ultraviolet zone
1800g /mm grating grooves, maximized resolution
Run in Windows operating system, the powerful software enables automatic qualitative and quantitative analysis. Automatic report generation in Excel format allows you to operate the instrument and process the data easily. You operation doest get any easier with our graphical interfaces.

Applications

Geology, minerals, metallurgy, steel, non-ferrous metals
Environmental analyses: air, water quality, soil and solid wastes
Petrochemical engineering, crude oil and related products, light industrial products,
Food, biomedicine and health products
Building materials (glass, ceramic, paints, etc)
Steel and non-ferrous metals analysis
Soil and solid wastes analysis
Biomedicine and health products analysis
Food analysis

Specifications

Model

AAS 6000

Optical system

all-reflective single-beam CT optical path

Focal length of monochromator

350mm, oil/water proof air compressor

Blazed wavelength of gratings 230 nm

Number of grating grooves

1800 g/mm

Wavelength range

190 nm - 900 nm

Wavelength accuracy

±0.1nm

Wavelength repeatability ±0.1nm
Spectrum bandwidth

0.1nm, 0.2nm, 0.4nm, 0.7nm, 1.4nm

Noise 0.005 Abs(static); 0.006 Abs(Dynamic)
Baseline drift 0.003 Abs/0.5h; best performance
Background correction D2 lamp + self reversal

Number of the lamps

8

Number of preload lamps software controlled, ≤8

Gas path safety measures

yes

Atomizer flame atomizer
Automation lamp / slit / wavelength / gas path / ignition / burner / safety measures / C2H2 monitoring
Standard configurations AAS6000
AAS6000 mainframe
Oil-free air compressor with lower noise
Special pressure gauge and peripheral accessories
Hollow cathode lamp
PC (brand computer, P4, LCD) 1 set
Printer (Canon, colorful ink-jet printer) 1 set
Wooden case packing and transportation

This specification can be changed without prior notice.