EDXRF Spectrometers |
|
|
|
|
|
X-CITE Series |
|
| Features |
The X-Cite system utilizes 35kV, 9-watt X-ray technology to analyze samples without damaging them. This approach is particularly valuable in applications where limited sample quantities are available or where sample integrity must be preserved for other purposes. |
| Applications | X-Cite - S: S in Oils and Fuels; Pb in Gasoline; Ni and V in Crude Oil; S, Ni, V in Residual Oil; S in Coke or Carbon; Mg, P, S, Ca, Ba, Zn, Mo in Lubricating Oils X-Cite - F: The Forensics industry can identify elements found in glass, steel, gun shot residue, automotive paints, and items used in the manufacture of drugs. X-Cite - A: Was created as a teaching tool in a variety of scientific fields as a way to introduce students to the EDXRF technique The EX-Cite “A+” will prove a great asset to any chemistry, physics, chemical engineering, geological, earth science or materials science teaching resource. X-Cite - C: Conforms to ASTM method C-114 for use in the Cement industry Na, Mg, Al, Si, S, K, Ca, and Fe in Cement, Clinker, and Raw Meal X-Cite-PPCP: Perfect tool for identification of active ingredients in cosmetics and pharmeceuticals: Cr, Mn, Fe in eye shadows; Fe, Ti and Zn in base makeup; Bi, Al, Ca, Mg in antacids, laxatives, gastrointestinal ailment remedies; personal hygiene products (deodorants, hair care, dental hygiene); antibacterial agents, Ag, I and Zn in fabrics, contact lenses; Ti and Zn in sunscreen. X-Cite-M: Designed for measurements of major elements and major impurities in: Zn metal, Zn powder, ZnO production; Fe metal, Fe powder, Fe2O3 production; Bauxite mining –aluminum; Kaolin clay analysis – ceramics; Tin ores; Silver ores |
|
Specifications |
|
| Excitation type | Direct (polychromatic) |
| Detector Type | New Generation PIN Diode |
| Tube Type |
Side Window |
| Maximum tube potential | 35 kV, 1 mA |
| Maximum tube power | 9 Watts |
| Elements analyzed | Na-U |
| Number of tube filters | 6 software selectable filters; open, Ti, Cu, Fe, Rh, W Filters can be optimized per application |
| Standard autosampler | 1samples (42mm max. diameter) and Large Sample chamber. Optional 8/16 position sample chamber |
| Computer | Integrated Data System high Resolution Color Printer |
| Software | nEXt Windows based analytical software |
| Power requirements | 115 or 230 VAC (1,2 or 0,6A respectively), 50/60Hz |
| Standard detector resolution | 160eV ± 10eV at 5,9keV |
| Data processing and storage | Advanced combined Analog-Digital processor with gated base line restorer, digital zero correction with integral multichannel analyzer and digital peak detector |
| Shock testing certifications | Meets MIL-Spec810E requirements (can be installed and operated in vehicle) |
| Temperature control | ± 5ºC at 25ºC |
| Storage temperature | up to 50ºC for 2 hours |
| Dimensions | 55×55×32 cm (unpacked) 75x80x62 cm (packed) |
| Weight | 80 kg(gr. weight) / 40 kg (net.weight) |
|
This specification can be changed without prior notice
|
|