Specifications |
| General Specification | Wavelength range | 190-900nm |
| Wavelength accuracy | ±0.5nm |
| Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30% |
| Baseline stability | 0.005A/30min |
| Background correction | The D2 lamp background correction capability at 1A is better than 30 times |
| Light Source System | Lamp turret | 2 lamps are powered simultaneously (one preheated) |
| Lamp current adjustment |
0~20mA |
| Lamp power supply mode | 400Hz square wave pulse |
| Opitical system | Monochromator | Single beam, Czerny-Turner design grating monochromator |
| Grating | 1800l/mm |
| Focal length | 277mm |
| Blazed wavelength | 250nm |
| Spectral bandwidth | 0.1nm, 0.2nm, 0.4nm, 1.2nm, automatic change |
| Flame Atomizer | Burner | 10 cm single slot all-titanium burner |
| Spray chamber | Corrosion resistant all-plastic spray chamber |
| Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate:6-7ml/min |
| Position adjustment | Manual adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner/min |
| Gas line protection | Fuel gas leakage alarm/min |
| Detection and Data Processing System | Detector | R928 Photomultiplier with high sensitivity and wide spectral range. |
| Electronic and microcomputer system | Automatic adjustment of light source power. Light energy and negative high-voltage auto-balance |
| Display mode | LEC display of energy and measurement values, concentration direct reading. |
| Read mode | Transient, time average, peak height, peak area, integral time is selectable in the range of 0.1~19.9s. |
| Scale expansion | 0.1~99 |
| Data processing mode | Automatic calculation of mean, standard deviation and relative standard deviation. Reporting number is in the range of 1~99 |
| Measurement mode | Automatic curve fitting with 3~7 standards; Sensitivity auto-correction |
| Result printing | Measurement data, working curve, signal profile and analytical conditions can all be printed out |
| Instrument self-check | Check current status of each function key |
| Characteristic Concentration and Detection Limit | Air-Acetylene | Cu: Characteristic concentration ≤ 0.25 mg/L. Detection limit ≤ 0.006 mg/L |
| Function expansion | | Hydride vapor generator can be connected for hydride analysis |
| Dimensions and weight | Main unit | 102x49x54 cm; unpacked 80kg |
| |
This specification can be changed without prior notice
|